12 Diamond X-ray tube data sheets
90 GE Healthcare 86500-IMG rev 3
12.8 FILAMENTS
12.9 STATOR 50/60 HZ
PHASE SHIFT CAPACITOR: 30 Microfarad
Safety: Thermal Switch
Connected in series with stator common wire
Opens at 79.4°C ± 3.9°C (175°F ± 7°F)
Closes at 68.3°C ± 3.9°C (155°F ± 7°F)
Low Voltage Cable:
Pin Position Function
1Phase Shift
2Phase
3 Common
4 Housing Ground
5Grid
6Not used
7 Small Filament
8 Filament Common
9 Large Filament
Focal spot Volts Amperes
0.1 3.8 - 7.3 3.0 - 4.8
0.3 6.0 - 10.5 3.6 - 5.2
Starting 115 V 230 V
Amperes 4 8
Seconds to
start
10.5
Duty Intermittent Intermittent
Running 45 V 90 V
Amperes 1.4 - 2.0 1.4 - 2.0
Duty Continuous Continuous