10 Specifications
86500-IMG rev 3 GE Healthcare 71
10Specifications
System specifications for Diamond, model MGX 2000
Manufacturer GE Healthcare Finland Oy
Kuortaneenkatu 2, FIN-00510
Helsinki, Finland
Manufacturer's quality system
according to
ISO9001
Manufacturer's environmental
management system according
to
ISO14001
General Medical electrical equipment
Model MGX 2000
Power supply Mains plug connection, 1-phase
(line, neutral and protective
ground, or L1, L2 and protective
ground)
Apparent resistance of supply
mains
0.75Ω
Rated voltage 230 VAC ±10%, single phase, 50/
60 Hz
16 A slow blow fuse in power
outlet
NOTE!
The Diamond is a cord-connected device, delivered without a plug (in
some markets). Depending upon local electrical code requirements, it
may be either hard-wired or plugged into an outlet. Please consult your
electrician to determine the requirements, and if necessary to provide a
plug which matches your outlet.
Line Regulation: ± 2% (full load)
Branch Circuit: dedicated run 20A, 2-pole circuit
breaker L1/L2 (10 AWG)
Outlet per local electrical code
requirements or NEMA L6-20
Grounding: 0V potential (earth), continuous
wire run (not conduit)
Rated current 20 A Momentary (during exposure)
300 mA long term (otherwise)
Rated input power 4.6 kVA
IEC safety class I