9 System maintenance
70 GE Healthcare 86500-IMG rev 3
9.6 REPLACEMENT OF THE LIGHT FIELD BULB
1 Turn the unit off and leave it for a while to allow the bulb to cool
down.
2 Slide the lamp support plate open.
3 When you see the old bulb lamp, pull it out.
4 Insert a new bulb. Avoid touching the bulb with your fingers. Use a
protective plastic bag (around the bulb, dependent on the bulb
manufacturer) or cotton gloves.
5 Slide the lamp support plate back to normal position.
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).