10 Specifications
86500-IMG rev 3 GE Healthcare 77
The nominal tube current 50 mA is obtained at the highest tube
voltage 39kV.
The highest tube current 100 mA is obtained at 25 kV tube voltage.
The highest electric power is obtained at 32kV tube voltage and 80 mA
tube current.
The nominal electric power of 2.4 kW is obtained, when loading time is
0.1s, tube current at 80 mA and x-ray tube voltage at 30 kV.
Fuses
Power Board 85020, F1,2 Fuse 5x20 mm 4AT, IEC 127-2/3
F3 Fuse 6.3x32 mm, 15ASB 250V
F4 Fuse 5x20 mm 0.8AT, IEC127-2/3
F5 Fuse 13/32” 20AF 600VAC
(Bussmann KTK-20)
AMD 32305, F1 Fuse 6.3x32 mm 10AT 250V
Diamond fuses used by Delta 32
On PCB located in the bottom-part plate of the C-Arm.
Fuse Board 33615, F5 Fuse 1 pc 6x32 mm 1,25AT 250V
UL
F3 Fuse 1 pc 6x32 mm 0,8AT 250V UL
F1,2,4 Fuse 3 pcs 6x32 mm 0,5AT 250V
UL
In the carriage next to the transformers: D32 Fuse Plate 82685
F1,2 Fuse 2 pcs, 6X32 mm 0,3AD 250
UL 248-14
FOR TRAINING PURPOSES ONLY!