10 Specifications
86500-IMG rev 3 GE Healthcare 79
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
IEC 61000-4-11
0 % U
n
for 5 sec 0 % U
n
for 5 sec Mains power quality is that of a
typical commercial and/or hospital
environment. If the user of the
Diamond requires continued
operation during power mains
interruptions, it is recommended that
the Diamond be powered from an
uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields are
at levels characteristic of a typical
location in a typical commercial and/
or hospital environment.
NOTE: U
n
is the a.c. mains voltage prior to application of the test level.
The Diamond is suitable for use in the specified electromagnetic environment. The purchaser or user
of the Diamond should assure that it is used in an electromagnetic environment as described below:
Immunity Test IEC 60601-1-2
Test Level
Compliance
Level
Electromagnetic Environment
Table 10.2 Electromagnetic immunity IEC 60601-1-2