General Output Rating and Duty Cycle:
At 230 VAC nominal:
Large Focal Spot 100 mA at 25 kV, 70 mA at 35 kV
Small Focal Spot 30 mA at 23 kV, 20 mA at 35 kV
Exposures are automatically limited if the tube or housing heat capacity
counter is activated.
Tube protection counter:
Anode:
Heating = kV x mAs = Joule (increased only during exposures)
Cooling = 30 Joule / sec. (decreased continuously)
Housing:
Heating = kV x mAs = Joule (increased only during exposures)
Cooling = (decreased continuously) refer to the diagram in Chapter 12.3,
Housing heating and cooling curves