Electrical Requirements 5.6 Typical Customer Supplied Cables and Wiring
5.6.1 Power Input Distribution Box (PIDB) with MDP (A1)
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare 5-16
Table 5-6: D670 Customer Supplied Cables and Wiring
Wiring
Description
Maximum run length
Meter (ft.)
Quantity Size AWG
(mm
2
)
From To
Run No. 1 – from primary power source to MDP
*
(see Figure 5-5,p.5-17)
3
*
* Refer to Table 5-3, D670 Minimum Feeder Wire Size,p.5-4 for AWG (mm2) wire sizes
Power 1 (3) 1 (3)
1 1/0 (50) Ground 1 (3) 1 (3)
Run No. 2 – from MDP to PDU * (see Figure 5-5, p.5-17)
3
*
Power 1 (3) 1 (3)
1 1/0 (50) Ground 1 (3) 1 (3)
1
*
Neutral (not required) 1 (3) 1 (3)
Run No. 3 – from MDP to System Emergency Off (SEO) (see Figure 5-5, p.5-17)
2 14 (2) Power 2 (6) 2 (6)
1 14 (2) Ground 2 (6) 2 (6)
Run No. 4 – CT PDU to warning light control (see Figure 5-6, p.5-18)
2 14 (2) Warning light 24 volt control A3J2-1,2,3,4 N/A N/A
Run No. 5 – CT PDU to scan room door interlock (see Figure 5-6, p.5-18)
2 14 (2) Scan room door interlock A3J6-1,2 N/A N/A