General System Requirements 1.3 Delivery Requirements
1.3.1 General Transportation and Delivery Precautions
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare 1-5
1.3.1 General Transportation and Delivery Precautions
1.3.1.1 General Temperature Precautions
Extreme temperatures must be avoided during system transportation and delivery. Ensure that
the system is not exposed, for an extended period of time, to temperatures or humidity outside
the following specifications.
Temperature range: -34ºC to +60ºC (-29ºF/+140ºF)
Humidity range (non-condensing): 5% to 95%
NOTE
Component freezing occurs if the system is exposed to temperatures below -18ºC (0ºF) for a period of longer
than two days. Allow a minimum of 12 hours for the system to adjust to ambient room temperature, prior to
installation.