Equipment Description and General Construction Requirements 2.3 Room Structural Requirements
2.3.1 Floor Requirements
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare 2-28
2.3.1.2 Floor Loading Requirements
Table 2-2: Weight of Components
Component
Weight (kg) Weight
(lb)
Load Distribution Comments
NM gantry (with HEGP
collimators mounted on
system)
2190 4828 4 pads, Ø83 mm each:
+845 kg each on front pads
+250 kg each on rear pads
Patient table
565 861246 4 wheels + axis anchored
to floor
Weight of table without
patient
Storage cabinet, including
tools (optional)
< 130 < 287
4 legs
CT gantry 1850 4078
CT PDU
D670/BSE 350
D670/OPT 370
D670/BSE 772
D670/OPT 815
4 wheels; anchored to floor
CT NIO console 89 196 4 wheels
AW workstation (optional) 30 66 Placed on desk
Collimator cart (with 2 HEGPs
on cart)
330 728 4 wheels
COG point at 99 cm height;
may be more than one
collimator cart in scan room
NM Acquisition station, ECG
Trigger Monitor, Operator
chair, Contrast Injector
(insignificant)
Personnel and patient
< 500 < 1102
Variable
Normally 3-4 people in room
during scan/service operations