Electrical Requirements 5.7 Lighting Specifications
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare 5-18
5.7 Lighting Specifications
Figure 5-6:D670 Typical TS6 Warning Light and Door Interlock Connections
10
9
Door Switch
PGND
EXP_INTLK
Run # 4
1
2
34
N.O.
Door Switch
5
6
78910
(Terminal TS6 in PDU)
Run # 5
Note:
Warning light and door
interlock are not mandatory in
some countries. If you install
an x-ray warning light in the
room, it is recommended that
you use the 4-wire method
which:
Minimizes EMC interference
Increases contact life of the
relay used in the PDU
If a door switch will not be
used, an additional jumper
must be added to TS6 in the
PDU during installation.
If the jumper is not in place,
exposures will not be made.