Equipment Description and General Construction Requirements 2.2 Room Size, Layout and Considerations
2.2.2 System Layout Drawings
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare 2-15
Figure 2-9: D670/OPT Reduced Serviceability Configuration Room Layout
Defines the minimum area
required to enable installation,
operation and service of the
system in safe conditions.
Does not take into account
local requirements. If local
requirements are stricter, they
should be followed (see
App.D, Regulatory Clearances).
Operator movement around
the system is limited.
Cannot use table extender.
Does not take into account
space for the following:
Operator room
Collimator carts
Optional UPS
No option for permanent
objects around the CT (for
example: wall ducts, sink,
cabinet.)
Requires Small Room
CT Gantry Installation Dolly
during installation.
Note:
Dimensions are in mm (")