1-19Girae Warmer
Temperature regulation alarms
Baby Hot-Check Temp Probe
Screen
Step
Description
1
Baby Mode
The “Baby Hot- Check Temp Probe” alarm
activates when the baby’s skin temperature
is greater than 1.0°C above the Set Temper-
ature. Silencing the alarm suspends the low
priority audio alarm for twelve (12) minutes.
When the baby’s temperature is greater
than 2.0°C above the Set Temperature a high
priority alarm activates. Silencing the alarm
suspends the high priority audio alarm for
ve (5) minutes.
The low priority alarm setting can be nar-
rowed to 0.5°C using the “Setup” screen and
“Pt Temp Alarm” keys.
Manual Mode
The alarm can be activated in the Manual
Mode using the “Setup” and “Man Temp
Alarm” keys. (for a detailed description, refer
to chapter 3).
Action: monitor the baby carefully and
check the attachment of the probe on the
baby’s skin.