EasyManua.ls Logo

GE Healthcare Brivo XR118 - Page 157

GE Healthcare Brivo XR118
318 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE
DIRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Chapter 6 - - Calibration Page 157
3.) Visible WC and WW Values
a.) No-Burn Mode (PACS Test Image 1)
i.) The DICOM header of each PACS Test Image is always sent with WC/
WW=5226/2654.
ii.) The values displayed on the PACS side may appear differently. This is
acceptable according to DICOM standards. The values can show up as 8192/
16384, 5226/2654, or something else. The most common and preferred
implementation is to show values of 8192/16384 on PACS.
iii.) If the values do not appear as 8192/16384:
Check that the VOI-LUT is being enabled on PACS. The VOI-LUT can often
be enabled on PACS with a drop-down menu on the image, or through
hanging protocols.
This does not indicate that the PACS does not support VOI-LUTs. Support
for VOI-LUTs is strictly determined by the visual appearance of PACS Test
Images 1 and 6.
iv.) Record results in Table 6-2, line number 7
b.) Burn-on-Send Mode (PACS Test Image 6)
PACS displays 8192/16384 for every image.
Record results in Table 6-2, line number 8
4.) Image Saturation
Open the PACS Test Image 1 and check that the clinical chest area appears with no saturation
in the lung region or the abdomen region. Saturation appears as a completely flat dark or white
area. Record results in Table 6-2, line number 9.
5.) Reduced Dynamic Range with Burn-on-Send
The Burn-on-Send configuration causes a reduced dynamic range in the image when the user
changes the window-level of the image on PACS.
Note: The images appear identically when first viewed. The loss of dynamic range only occurs when
the user changes window-level.

Table of Contents

Related product manuals