GE HEALTHCARE
D
IRECTION 5461425-8EN, REVISION 6 BRIVO XR118 SERVICE MANUAL
Page 254 Chapter 12 - - Set Preferences
General
This section provides instructions for setting your system preferences for
network and printer connections. The Services Desktop and Shut Down functions
are also available from this screen. Refer to Chapter 9: General for more
information about Shut Down.
Network Connections
Network and printer connections are configured through the Utilities [ ]
- General screen. This screen allows qualified service personnel to define the
Digital Imaging and COmmunication in Medicine (DICOM) send destinations.
Network connections may be added, removed, or edited from this screen.
Follow this process to access the Network Connections screen.
1. On the Worklist screen, click utilities button [ ].
The System - General screen appears.
2. Click the [Edit] button of [Network Connections].
The Network Connections screen (Figure 12-3) appears.
Figure 12-3 Network Connections screen