GE HEALTHCARE OPTIMA XR200AMX/XR220AMX
D
IRECTION 5336113-1EN, REVISION 10 INSTALLATION MANUAL
Page 100 Section 4.12 Enabling Advanced Service mode
4. Press CTRL + SHIFT + F5 to open the Engineering User Interface.
5. On the Engineering User Interface, press the TERMINAL
icon.
6. At the command prompt, type:
mount /media/cdrecorder
Press ENTER
.
cd /magichome/xruser/config/restore/
Press ENTER
.
7. At the prompt, type:
chmod 777 restoreCustomIPLooks_F2toF3.sh
Press ENTER
.
8. At the prompt, type:
./restoreCustomIPLooks_F2toF3.sh
Press ENTER
. The script takes approximately 20 minutes to run. When completed, the system
returns to the prompt.
9. At the prompt, type:
umount /media/cdrecorder
Press ENTER
.
10. Remove the Definium AMX 700 backup media from the CD/DVD drive and disconnect the
drive from the system.
11. Type exit to close the terminal window.
12. Exit the Engineering User Interface.
13. Reset the system.
Section 4.12 Enabling Advanced Service mode
Advanced Service (sometimes called “in-house service” or “extended service”) is available to
customers that have purchased an Advanced Service package license from GE Healthcare.
Contact a GE service representative to enable Advanced Service.
Advanced Service provides the following functionality:
• SUIF in Advanced Service mode
• Detailed error descriptions in the Service Error Log Viewer
• Advanced diagnostics
• Extended error log with filtering
The following functionality is never included in an Advanced Service license:
• InSite ExC configuration
• Advanced service configuration or entitlement
• Purchased options configuration
Section 4.13 IP Address Change Procedure
The system uses IP Addresses 192.168.x.x for communication between the PC and the detectors.
If the hospital network to the PC is also 192.168.x.x, issues will occur. Perform the following