EasyManua.ls Logo

GE Healthcare Optima XR200amx - Site Configuration

GE Healthcare Optima XR200amx
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARE OPTIMA XR200AMX/XR220AMX
D
IRECTION 5336113-1EN, REVISION 10 INSTALLATION MANUAL
Page 58 Section 4.2 Using the System Configuration Tool (SUIF)
4. On the Utilities desktop, next to Service, press LAUNCH.
The Service Home tab is displayed. The Home tab contains summary information including
site identification, connectivity parameters, and enabled options.
5. Press the CONFIGURATION
tab button.
6. On the Configuration page, in the left folder tree, select the parameter group to change.
7. Configure the system according to site preferences.
An overview of the configurable system attributes is given in the tables below.
Contact a GE service representative for items marked “Editable by GE Healthcare personnel
only.”
8. On each page, after making changes, press COMMIT
to put the changes into effect.
4.2.1 Site configuration
The following information can be viewed regardless of service mode. However, some fields are
editable only by sites with Advanced Service mode enabled (Section 4.12 Enabling Advanced
Service mode on page 100), and some are editable only by GE Healthcare personnel.
Table 4-2 Site tab parameters
Parameter Definition
Site Name Name of specific site where system is installed.
Street Address 1 Street address of site where system is installed.
Street Address 2 Street address, continued, of site where system is installed.
City City where system is installed.
State/Province State or province where system is installed.
Postal Code Postal or ZIP code where system is installed.
Department Name
Department name (for example, Radiology) responsible for system
maintenance.
Telephone Phone number of department responsible for system.
Fax Fax number of department responsible for system.
Contact Person Name Person responsible for system maintenance.
Contact Person Phone Phone number of person responsible for system maintenance.
System ID
Editable by GE Healthcare personnel only.
System identifier assigned by GE Healthcare following system
installation and prior to customer turnover.
This field is a GE proprietary field and cannot be used for any purpose
other than the System ID. Violation of this mandate will result in Insite
and/or other features not working. Customer IT personnel need to use
other means for DICOM filtering on their DICOM machines.
Installation Date
Editable by GE Healthcare personnel only.
Date system installation is completed, created by GE Healthcare.
Service Contract Number
Editable by GE Healthcare personnel only.
Customer’s Service Contract number, created by GE Healthcare.
Contract Expiration
Editable by GE Healthcare personnel only.
Customer’s Service Contract expiration date, created by GE
Healthcare.