Page 66 Section 4.2 Using the System Configuration Tool (SUIF)
4.2.10 Time Server configuration
Contact a GE service representative to configure Time Server settings.
4.2.11 Secure Login configuration
Select SAVE to apply your changes, or CANCEL to cancel them.
Use the arrow buttons to move through the pages.
Add New Filter
In the Name field, specify the filter name.
In the Allowed IPs field, specify a particular IP address, a range
of IP addresses, or a masked subnet.
In the Allowed Port field, specify a particular port.
From the Protocol drop-down list, select the protocol to be
allowed (TCP, UDP, Any).
Select ADD FILTER
to apply your changes.
NAT tab
Used to define NAT (Network Address Translation) rules.
Name/Source IP/Dest. IP
Lists current filters by name, source IP, destination IP,
destination port, and protocol.
Remove Existing Filters
Select the Delete box checkbox next to the filter to delete, then
select DELETE EXISTING FILTERS
.
Add New Filter
In the Name field, specify the filter name.
In the Source IP field, specify the source IP.
In the Destination IP field, specify the destination IP.
In the Destination Port field, specify the destination port.
From the Protocol drop-down list, select the protocol to be
allowed (TCP, UDP, Any).
Select ADD FILTER
to apply your changes.
GE Service tab
Used to set IIP node-lock IP address and DICOM port. This page is limited to GE service
personnel.
Table 4-10 PNF parameters
Option Description
Table 4-11 Secure Login options
Option Description
Audit Message Settings tab
Audit Source ID IP address of audit message source.
Patient Name
Anonymized
On or Off.
Enterprise Repository tab
Enterprise Repository 1 On or Off.