GE HEALTHCARE OPTIMA XR200AMX/XR220AMX
D
IRECTION 5336113-1EN, REVISION 10 INSTALLATION MANUAL
Page 78 Section 4.10 Installing hardware options
Figure 4-5 OPI-4002 barcode reader charging cradle and power adapters
4.10.2 Configuring the barcode reader
Applies to Optima XR220amx or Optima XR200 with digital upgrade systems only
The following section provides instructions for the default configuration and two optional settings.
Have a printed copy of these pages on hand so you can scan the necessary barcodes. Refer to the
OPI-4002 User’s Guide for additional configuration options.
4.10.2.1 Setting to default configuration
1. Turn on the system. Wait until the green LED on the barcode reader receiver starts flashing.
2. Press the barcode reader trigger to start the scan.
3. Scan the Reset To Default barcode.
Figure 4-6 Reset to Default barcode
You will hear three quick sets of high-low-high beeps.
Whenever you scan the Reset To Default barcode, rescan the barcode label (on the barcode
reader receiver) and re-program any optional configuration.
4. Scan the barcode label (Figure 4-1) on the barcode reader receiver to register the barcode
reader.
You will hear one set of high-low beeps, a short pause, and then one long high beep.
The connection is complete and the barcode reader is ready for verification.