Page 86 Section 4.10 Installing hardware options
holder assembly on the column.Apply thread locker (i.e. Loctite 243 ) to the 4 tablet holder
assembly mounting screws (new item 1), and tighten the screws. Torque to 3.2 N-m (2.36 lbft).
Figure 4-15 Installing the tablet holder assembly
5. Unlock the tablet clamp assembly with the key.
6. Lift up the top cover of the tablet holder clamp.
Item Description
1 Tablet holder assembly mounting screws (4)
2 Tablet holder assembly
3 Lead apron hanger