Illustration 5-15: Floor Space Diagram for Lytron Chiller
The maximum elevation difference between X–Ray Tube Housing and Chiller is
around 3 meters (10 ft).
Exact dimensions of the Coolix 4000 chiller depend on model type. Required floor space depends
on ambient room temperatures. When in doubt, allow for maximum floor space.
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2
126 1 Dimension Drawings