6 Configuration Checks Before Innova IQ Table Upgrade
1.
Check that the Omega table is installed on a base plate. If not, order a base plate for the
Innova
IQ
Table and plan flushing it in the concrete as part of the upgrade.
2.
Check that the Omega base plate is flushed in the concrete If not, plan flushing it in the concrete
as part of the upgrade. If not possible contact engineering.
3.
Check if the base plate is drilled for Innova
IQ
Table anchorage. If not, plan to rework the base
plate as part of the upgrade.
4. Check that it is possible to pull the cables connected to the table (1 meter) out of the cable
channel. If it is not possible, order a new cable set of appropriate length and install this new
cable set as part of the upgrade.
5. Check that the system software configuration is Unity SP3.2 or higher (DL application revision
is DL_Unity_4.18.9 or higher). If not, plan to upgrade the site with the relevant upgrade and /
or FMI
6. Check that clearance with floor, walls and ceiling will comply with room layout and physical
characteristics.
7.
If the customer uses a Smart Handle with the Omega, check that he/she has been notified that
the Smart Handle will be replaced by a Smart Box (there is no Smart Handle with the
Innova
IQ
Table).
8. One TSSC and one Smart Box are supplied with the Upgrade Kit. If the site configuration
requires more, check that the customer order contains a remote box (P/N 5184336) and the
correct number of extra TSUI (TSSC and Smart Box).
9. If the customer uses an injector head table mounted or an entire injector rail mounted, check
that he/she has been notified that this injector head cannot be mountable on the Innova
IQ
Table and that an alternate solution has been implemented (using a pedestal, for instance).
10. If the customer uses a table top rear support with the Omega, check that he/she has been
notified that this accessory cannot be used with the Innova
IQ
Table.
11. Write down the DL Name which will be needed later on during software installation. To find it,
right click on the icon
My Computer
, select [Properties] and go to the
Network identification
tab. The DL name is displayed in the
Full computer name
field.
12.
Innova
IQ
Table does not support the connectivity for IVUS and MacLab. Ensure that IVUS and
MacLab cables are routed through a dedicated conduit.
13.
Case of CA1000: the Innova
IQ
Table is not compatible with CA1000 workstations version prior
to SPa07. If the site is using a CA1000, check the current version and upgrade it to SPa07 or
higher before installing the upgrade.
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2
Chapter 7 Additional Planning Aids 251