1.3 Description of the X–Ray generator
The Innova System uses an Innova High Voltage System with component parts as follows
according to the power requested (100 kW):
Jedi – C1 cabinet,
Ingrid H.V. tank mounted on the Performix 160A X–Ray tube housing.
1.4 Description of the X–Ray head
The Innova System uses a Performix 160A X–Ray tube housing:
X–Ray tube
collimator
Ingrid HV tank
oil/water exchanger
contour filter
tube chiller
NOTE: An external recirculating chiller is mandatory.
It is mandatory to place the Chiller in the Technical Room.
1.5 Description of the Innova Imaging System
An Innova System is managed and controlled by a System including (does not list all components):
C1 cabinet, inside which the main components are the RTAC, the DL, the modem, the Firewall,
the HUB, the KVM, the 4 KVA PDU and the Jedi generator,
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2
30 1 Description of Innova System