2.1 Room layout and room min length
Table 4-1: Exam Room dimensions
Length
Width
Ceiling Height
Recommended:
32 ft 0 in (9.75m)
20 ft 0 in ( 6.00 m)
10 ft 0 in (3.05 m)
Minimum:
For minimum exam room lengths, see different table configu‐
rations in Illustration 4-2, Illustration 4-3 and Illustration 4-4
14 ft 5 in (4.40 m)
9 ft 0 in (2.74 m)
Note (1): With a 3 monitor frame suspension, take care with the ceiling height not more 3.05 m (10 ft)
Illustration 4-2: Omega V or Innova
IQ
Table (Angio, Cardio, Neuro) with header extender
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2
90 2 Room Layout Drawings