2.3.3.4 2285052 – Above grade, Floor anchor with insert
Table 5-11:
Item
Name
Part #
Description
Quan.
Notes
Floor Anchor
46-302265P1
Floor Anchor Bolt; 5/8-10 x 6 in. drill‐
ing
12 anchors for LC Positioner and
4 anchors for Table
16 pc
Above Grade anchor mount‐
ing method hardware
Grout assy
2285055
- 10 kg Powdered Mortar Ardex
K-15
- RTV Silicon Rubber Adhesive
- 18 mm making tape adhesive
- Open cell foam
1 kit
Used in constructing LC Po‐
sitioner grout dam
Dowel
2290937-2
Wood Dowel; 16 mm diameter
12 pc
Water Elec‐
tric Separator
2268647
Manufactured part introduced in Po‐
sitioner key hole to fit 1 electrical
conduit of 6”.
1 pc
Electrical and water conduits
separated inch size.
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2
158 2 Mounting Requirements

Table of Contents