13 Advantage Windows
A power cord is supplied with Advantage Windows. It is connected to outlet fed by the main
disconnect room device.
Verify that the ground between AW, AW spliter and monitors ground are equipotential (0.1 ohm
equipotentiallity between room ground point (busbar) and wall outlets).
To power the AW splitter, use the appropiate cable out of lot P/N 2385173.
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2
Chapter 6 Electrical Connections 197