IV apparatus.
Injectors.
Heart monitoring equipment.
Crash cart.
1.5 Emergency Stop
Protect the Emergency Stop from accidental actuation.
1.6 Patient Environment Equipment
The components that may be installed within patient vicinity need to be medical equipment (”patient
vicinity is defined in the standardization as a space within the room 1.83 m (70.7 in) beyond the
perimeter of the examination table and extending vertically 2.29 m (90.2 in) above the floor.”). For
the Innova System, the equipment are:
Table
C–arm
Monitors
Injector
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2