2.4 Injector Mounting Requirements
Table accessory rail load consideration:
The maximum load per table accessory rail is 40 kg (88 lbs) at 150 mm (0.49 ft) (60
N.m or 44.25 ft/lbs). Therefore:
Only light extra load not exceeding 5 kg (11 lbs) at 100 mm (0.33 ft) (i.e IV pole
with its accessories, pressure head...) is authorized on the same table
accessory rail as the injector.
Never install injector and radiation protection on the same table accessory rail.
Typical installation on the front table accessory rail is Smart handle or Smart
box, Table Side System Control (TSSC), InnovaCentral/Touchscreen, Table
panning device and cables support.
If needed an optional rail can be installed at table foot end of the Omega V table
for other options.
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2
164 2 Mounting Requirements