VFI (Voltage Frequency Independent) UPS system where the load is continously supplied by
the inverter through the rectifier.
SEM (Super Eco Mode) permitting the maximum energy saving.
Table 3-4: Fluoro UPS CE
Losses
Cooling air flow
VFI
SEM
VFI
SEM
2.14 kW
0.64 kW
625 m
3
/h
190 m
3
/h
Table 3-5: Fluoro UPS UL
Losses
Cooling air flow
BTU / hr
kW
CFM
m
3
/h
6751
1.98
301
512
If installed in the technical room, the UPS may impact on it’s layout. It may also be
installed in a separate room. This depends on hospital constraints, local regulations
or EHS rules. Clearance, weight of UPS, airflow and cooling system should be
adapted for the UPS.
ELECTRICAL CONTRACTOR IS RESPONSIBLE FOR PROVIDING AND
CONNECTING THE CABLES FROM THE PDB TO THE UPS AND
CONFIGURING THE PDB IN BYPASS MODE. GEHC IS RESPONSIBLE FOR
POWERING ON THE SYSTEM WITH THE UPS IN BYPASS MODE. GEDE IS
RESPONSIBLE FOR UPS COMMISSIONING.
The Fluoro UPS can be installed either in an Electrical Local with restricted access
or in the innova technical room as per local regulations.
Make sure that local regulations have been applied for the installation of the Fluoro
UPS (dedicated room/fire detection etc.)
NOTE: Refer to the UPS vendor Service manual
Installation Guide SG Serie 10, 20, 30 & 40
kVA for more details
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2