Illustration 3-7: Water Conduit location
In some countries, depending on local regulations, it may be forbidden to run
electrical cables and water pipes in the same conduit. In this case, two separate
conduits are required.
Innova 2100-IQ, 3100/3100-IQ, 4100/4100-IQ Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare Direction 5160944-13-1EN, Revision 2
72 3 Structural Requirements