IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
4 - 2 Section 4-3 - General procedures
Section 4-3
General procedures
CAUTION
!! CAUTION:
SYSTEM REQUIRES ALL COVERS
Operate this unit only when all board covers and frame panels are securely in place.
The covers are required for safe operation, good system performance and cooling purposes.
NOTICE
Energy Control and Power Lockout for LOGIQ S8.
When servicing parts of the system where there is exposure to voltage greater than 30 Volts:
1. Complete shutdown the system.
2. Unplug the system.
3. Maintain control of the system power plug.
4. Wait for at least 20 seconds for capacitors to discharge as there are no test points to verify isolation.
Beware that the Power Supply, Front End Processor and Back End Processor may be energized even
if the power is turned off when the cord is still plugged into the AC Outlet.
UT

Table of Contents

Other manuals for GE LOGIQ S8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S8 and is the answer not in the manual?

GE LOGIQ S8 Specifications

General IconGeneral
TypeUltrasound System
Display TypeLCD
3D/4D ImagingYes
ConnectivityDICOM, USB, Ethernet
Display Size19 inches
TransducersLinear, Convex, Phased Array
Doppler ModesColor Doppler, Power Doppler, Spectral Doppler
Operating ModesB-Mode, M-Mode
ApplicationsAbdominal, Cardiac, Obstetrics, Gynecology, Vascular, Musculoskeletal
PortabilityCart-based
Transducer Ports4

Related product manuals