IRECTION 5394227, 12 LOGIQ S8/LOGIQ E8 SERVICE MANUAL
Section 4-3 - General procedures 4 - 17
4-3-8-4 Setting up at a new location
1.) When the unit is in place at a new location, lock the wheel brakes.
2.) Follow the set up procedure described in Section 3-6 "Completing the Set Up" on page 3-12.
4-3-9 Service Platform Confirmation (R1 to R4)
1.) From the touch panel, select Utility -> Service.
2.) On Service Browser, select GE Service, and enter the password (See Section 7-6 "Service Desktop
(CSD) - R1 to R4" on page 7-24 for details).
3.) Note “SvcPform” version.
Figure 4-14 Service Platform version

Table of Contents

Other manuals for GE LOGIQ S8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S8 and is the answer not in the manual?

GE LOGIQ S8 Specifications

General IconGeneral
TypeUltrasound System
Display TypeLCD
3D/4D ImagingYes
ConnectivityDICOM, USB, Ethernet
Display Size19 inches
TransducersLinear, Convex, Phased Array
Doppler ModesColor Doppler, Power Doppler, Spectral Doppler
Operating ModesB-Mode, M-Mode
ApplicationsAbdominal, Cardiac, Obstetrics, Gynecology, Vascular, Musculoskeletal
PortabilityCart-based
Transducer Ports4