EasyManua.ls Logo

GE Sievers 500 RL - Accuracy, Precision, and Verification Worksheet

GE Sievers 500 RL
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE Analytical Instruments ©2010 136 of 226 DLM 74001-04 Rev. A
Accuracy, Precision, and Verification Worksheet
Company Name ____________________________ Date _________________________
Analyst Name ______________________________ Firmware Version ______________
Analyzer Serial Number ______________________ Standards Expiration Date _______
Standards Set Lot No. (optional)________________ Standards Vials Lot Nos._________
Rep RW TOC 500 ppb TOC 25.00 µS/cm tCond
1 ________ ________ ________
2 ________ ________ ________
3 ________ ________ ________
Avg ________ ________ ________
RW TOC Conductivity
Avg ________ ________ ________
Adj Standard
Concentration ________
N/A N/A
SD N/A ________ ________
RSD (verification)
N/A ________ ________ Pass Fail
Accuracy N/A ________ ________ Pass Fail
Pass criteria:
TOC Precision: RSD of last three measurements of 500 ppb standard
3%
Conductivity Precision: RSD of last three compensated conductivity measurements on 25 µS/cm
standard
2%
TOC Accuracy: % Difference
7%
Conductivity Accuracy: % Difference
2%

Table of Contents