Chapter 3: Installation
which is adjusted by the screw on the sample inlet system. Turn the screw clockwise to decrease flow,
and turn the screw counter-clockwise to increase flow.
Step 7: Fill the DI Water Cartridge
Fill the DI water cartridge with DI water, using the squeeze bottle provided in the accessories kit.
1. Fill the water bottle with DI water.
2. Open the Analyzer’s front panel (you will need the key from the accessories kit to open the bottom
latch).
3. Locate the DI water cartridge at the center of the Analyzer (see Figure 7)
Figure 7: Locating The DI Water Cartridge
4. Lift the spring clip (see Figure 8 on page 57) at the top of the DI water cartridge, and rotate the cartridge
clockwise (as viewed from the top) so it slides out of the retaining hooks.
Note:
Setting a water flow rate that exceeds 300 mL/min may result in poor system
suitability verification data or any measurements taken from sample water in vials.
Excessive flow also will not allow the sample pump to draw sample properly.
Note:
The iOS System door may be taped closed for safety during transport. Remove the
tape before proceeding.