GE Analytical Instruments ©2010 162 of 226 DLM 74001-04 Rev. A
Chapter 8: Troubleshooting
Figure 41: The Inlet and Outlet Connections Inside the Analyzer
9. Swap the two pairs of connections leading to the sample pump tubing at the plastic barbs (see
Figure 42). The first pair of tubes (on the left) should be connected to the barbs that normally connect to
the second pair of tubes (on the right). The second pair of tubes (on the right) should be connected to
the barbs that normally connect to the first pair of tubes (on the left).
Figure 42: The Sample Pump Tubing Connections