Chapter 7: Maintenance
Figure 32: Interior Overview of the Analyzer
Checking and Refilling the DI Water Level
The water level in the DI water cartridge should be checked periodically and refilled as necessary. For this
procedure, you will need the water bottle (from the accessories kit) filled with DI water.
To check the level of water in the DI water cartridge:
1. If the Analyzer is taking measurements, press the Stop Analysis button.
2. Turn off power to the Analyzer with the main power switch.
3. Open the Analyzer front panel.
4. Locate the DI water cartridge at the center of the Analyzer (see Figure 32 on page 140).
Table 13: Replacement Schedule for Consumables
Description Typical Operating Life
DI Water Level Refill as needed (check when replacing other items)
UV lamp 6 months
Pump tubing (with pump heads) 12 months
DI Water Cartridge Typically 24 Months (depends on water quality)
Inline filter Replace as needed (depends on water quality)