EasyManua.ls Logo

GE Sievers 500 RL - Page 40

GE Sievers 500 RL
186 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE Analytical Instruments ©2010 40 of 226 DLM 74001-04 Rev. A
Chapter 2: System Description
Conductivity
Raw Conductivity Range 0.01 - 35 µS/cm
Conductivity Accuracy ±0.005 µS/cm
0.25 µS/cm
±1%
0.25 µS/cm
Conductivity Precision <0.25% RSD
Maximum Sample Conductivity 25 µS/cm at neutral pH
Instrument
Power requirements 100-240 ±10% VAC, 100 watt, 50/60 Hz
Fuses No user-replaceable fuses
Normal Operating Environment Intended for indoor use only
Ambient Temperature 10 °C to 40 °C (50 °F to 104 °F)
Maximum Relative Humidity Up to 95%, noncondensing
Maximum Altitude 2,000 m (7,546 ft)
Inputs One isolated binary input
Outputs Serial (RS-232) port, USB, parallel printer port,
three 4–20 mA ports, four alarms ports, one
Ethernet port
Installation/Overvoltage Category II (protects against transients present in Category II
power)
Safety Certifications CE, ETL listed. Conforms to UL Std. 61010-1.
Certified to CSA C22.2 No. 61010-1.
Pollution Degree 2 (normally only non-conductive pollution)
Display Backlit Quarter-VGA touchscreen display
Size 41.9 cm height x 48.3 cm width x 27.4 cm depth
(16.5 in x 19.0 in x 10.8 in)
Weight 16.9 kg (37.2 lbs)
IP Rating IP 45
*Note: Stated analytical performance is achievable under controlled laboratory conditions that minimize operator and standards errors.
**iOS and Super iOS only. For more information, see page 96.

Table of Contents

Related product manuals