EasyManua.ls Logo

GE Versana Balance - Page 167

GE Versana Balance
431 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Connectivity setup
Versana Balance – Basic Service Manual System view 3-87
5808768-100 English Rev.9
3-9-4 Setup connection to a DICOM server(continued)
Table 3-29: Utility ->Connectivity ->Device screen
1Select MyComputer.
2 Assign an AE title to the Versana Balance. (AE stands for Application Entity. DICOM
services use this to identify the Versana Balance.) AE title is case-sensitive. This
title may contain the Computer Name from the TCP/IP page, if desired.
Note: It is NOT recommended to use the factory default. This is not prohibited, but
more than one system with the same AE title can cause confusion.
3 Edit Port Number if needed. 104 is typical. Save your changes and reboot the
system.