Replacement Procedures
8-10 System view Versana Balance – Basic Service Manual
5808768-100 English Rev.9
8-4-4 Backing up the Patient Archive and System Configurations
• Backup the Patient Archive and System Configurations.
For instructions, please see “Data Backup and Restore” in
the Basic User Manual/User Guide.
8-4-5 Restoring up the Patient Archive and System Configurations
NOTE: It is not suggested to manually delete the files in
D:\Idunn\target\resources\idunn\userdefs.
An error, or a power loss may occur.
Always backup the Patient Archive and the Presets (System
Configurations) before loading the software!
In order to complete a successful restore of the Patient
Database, as needed after a hard disk replacement, or if all the
content on the hard disk has been erased, the images must be
moved away from Versana Balance before doing backup of the
Patient Database.
Depending on the location set-up, either move the images to a
remote server or to removable media like DVD or CD discs.
To avoid not being able to connect to Local Archive,
connecivity.res and IPSave in
“D:\Idunn\target\resources\idunn\userdefs” should not be
deleted. If they are deleted, please rewirte the serial number.