General Procedures and Functional Checks
4-36 System view Versana Balance – Basic Service Manual
5808768-100 English Rev.9
4-2-21 Monitor
To clean the monitor face:
Use a soft, folded cloth. Gently wipe the monitor face.
DO NOT use a glass cleaner that has a hydrocarbon base (such
as Benzene, Methyl Alcohol or Methyl Ethyl Kentone) on
monitors with the filter (anti-glare shield). Hard rubbing will also
damage the filer.
NOTE: When cleaning the screen, make sure not to scratch the
monitor.
4-2-22 Operator Control Panel
To clean the operator control panel:
1. Moisten a soft, non-abrasive folded cloth with a mild,
general purpose, non-abrasive soap and water solution.
2. Wipe down operator control panel.
3. Use a cotton swab to clean around keys or controls. Use a
toothpick to remove solids from between keys and controls.
NOTE: When cleaning the operator control panel, make sure not to spill
or spray any liquid on the controls, into the system cabinet, or in
the probe connection receptacle.
NOTE: In case of SARS, use bleach, or Cidex in a normal diluted form
for cleaning/disinfecting the operator panel.
NOTE: DO NOT use T-spray or Sani Wipes on the control panel.
NOTE: DO NOT remove AN Key film when clean the control panel.
CAUTION
Before cleaning the control panel, make sure the key cap is
firmly in place.