Diagnostics/Troubleshooting
7-20 System view Versana Balance – Basic Service Manual
5808768-100 English Rev.9
7-5-5-3 System Status
Use System Status to view status information on the Versana
Essential. Specifically, do the following:
• View the hardware configuration file
• View temperatures for FRUs, fans, and a graphical
representation of the monitored temperatures
• View voltages for FRUs and AC
• View the amount of time the Versana Essential has been
running
• View the amount of time since the Versana Essential has
last been rebooted
The information on System Status is available to Class C and
Class M licenses.
To access System Status, navigate to Utility > Service > Home.
Figure 7-16. System Status
This table shows all the elements available on System Status
with descriptions.
Table 7-3: System Status
Element DESCRIPTION
Hardware
Configuration
Displays the contents of the hardware configuration file.
Temperature Displays the temperature monitoring of the Versana Essential as normal, warning,
or error using the standard colors and icons. Includes temperature out-of-spec
events including an overall count of event warnings and errors for the last 30 days
or the last status checkpoint whichever is sooner. Temperature displays
temperature a graph for the configured monitoring points.