Diagnostics/Troubleshooting
7-46 System view Versana Balance – Basic Service Manual
5808768-100 English Rev.9
7-6-3-6 Assessment Utility
7-6-3-6-1e-PAT
e-PAT is a probe assessment tool that evaluates each probe
element response scanning on the air. Probes have to be clean,
any gel residue will provide incorrect data. This test is an
assessment that is intended to be used comparatively during the
life of the probe to evaluate possible probe deterioration over
time.
NOTE: Probe diagnostics is available to Class C (Service Pro) license.
1. Check that the probe to be tested is thoroughly clean and
dry. Connect it to the probe port on the scanner, then hold
the probe in the air ready for testing.
NOTE: The probe assessment is only available for the probe
connected in the first probe port (the leftmost probe port).
Figure 7-40. The first probe port
NOTE: Prior to inserting the probe, ensure that the connector
locking handle is positioned to the unlock state.
NOTE: Ensure the probe is locked to the system before you start
the probe diagnostics.
CAUTION
DO NOT allow the probe head to hang free. Impact to the
probe head could result in irreparable damage.

Table of Contents