Care and Maintenance
10-14 System view Versana Balance – Basic Service Manual
5808768-100 English Rev.9
10-6-2-4 Cleaning
For general cleaning of the system, we recommended to use
non-abrasive soap and water solution.
Table 10-9: General Cleaning
Step Item Description
1. Console Remove the battery. Use a fluid detergent in warm water on a soft, damp cloth to
carefully wipe the entire system. Be careful not to get the cloth too wet so that
moisture does not enter the console.
2. Probe Holder Clean probe holders. (they may need to be soaked to remove excess gel).
3. Monitor Use a soft, non-abrasive folder cloth. Gently wipe the monitor face. DO NOT use
a glass cleaner that has a hydrocarbon base (such as Benzene, Methy Alcohol
or Methy Ethyl Ketone) on monitor with the filter (anti-glare shield).
Table 10-10: Compatible Chemicals for Cleaning
System
Cabinet
Operator
Control
Panel
Main
display
Probe
Holder
Cleaning agents
PDI Easy Screen Cleaning Wipe X X X X
Wet Wipes X X X X
70% isopropyl alcohol X X X X
Disinfectants
[1]
PDI Sani-Cloth Plus X X X X
Protex X X X X
Tristel Wipes X X X X
Sono Ultrasound Wipes X X X X
Clorox wipes X X X X
CaviWipes X X X X
PDI Super Sani-Cloth Plus X X X X
PDI Easy Screen Cleaning Wipe X X X X
[1]: Effective Disinfection is always a balance between safe inactivation of infectious agents and
undesirable side effects. Due to the generally uneven and irregular surface of Ultrasound consoles, a
comprehensive surface disinfection process cannot be recommended by the manufacturer.