General Ultrasound system requirements
Versana Balance – Basic Service Manual System view 2-7
5808768-100 English Rev.9
2-2-3-4 Site circuit breaker
It is recommended that the branch circuit breaker for the
Ultrasound system be readily accessible.
2-2-3-5 Site power outlets
A dedicated AC power outlet must be within reach of the
Ultrasound system without extension cords. Other outlets
adequate for the external peripherals, medical and test
equipment needed to support this Ultrasound system must also
be present within 1 m (3.2 ft.) of the Ultrasound system.
Electrical installation must meet all current local, state, and
national electrical codes.
2-2-3-6 Unit power plug
If the Ultrasound system arrives without a power plug, or with
the wrong plug, you must contact your GE dealer or the
installation engineer must supply what is locally required.
POWER OUTAGE MAY OCCURE.
The Versana Balance requires a dedicated single branch
circuit. To avoid circuit overload and possible loss of critical
care equipment, make sure you DO NOT have any other
equipment operating on the same circuit.