GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-67
iPWCW_CloseProcessing in RT_PWorCWFilterBlock::execute() returned error
code xx !! Action: restart the system, If after restart not ok then call technical support
restart the system
iPWCW_InitializeProcessing in RT_PWorCWFilterBlock::execute() returned error
code xx !! Action: restart the system, If after restart not ok then call technical support
restart the system
iSetVideoSource(eVideoIntern) function fails
Action: The system will restart itself by pressing OK
press OK
LineFilt Out Of Range press ok and save this user-setting once again
Loading DVD data..... Requested operation could not be executed!
Try again after DVD data is loaded.
Press ok, wait until loading process is finished and try
again
LP_KoefBlock: SamplePRF too big restart the system
M_Gain Out Of Range press ok and save this user-setting once again
M_Reject Out Of Range press ok and save this user-setting once again
MC Ensemble Out Of Range press ok and save this user-setting once again
MC_Balance Out Of Range press ok and save this user-setting once again
MC_BaseLinePos Out Of Range press ok and save this user-setting once again
MC_Dynamic Out Of Range press ok and save this user-setting once again
MC_FallSmooth Out Of Range press ok and save this user-setting once again
MC_Gain Out Of Range press ok and save this user-setting once again
MC_RiseSmooth Out Of Range press ok and save this user-setting once again
MCShotBuffer : not enough memory reserved" ". Action: restart the system restart the system
M-DynContrast Out Of Range press ok and save this user-setting once again
memory allocation error Action: restart the system restart the system
Memory allocation failed Action: restart system restart the system
MessageBox(NULL, err_msg, "CVIE ERROR", MB_OK); restart the system
missing BCMCPW_TxFocusData Action: restart the system restart the system
missing BM_TxFocusData Action: restart the system restart the system
missing ProbeAcousticUnit, wrong ProbeAcousticUnitID. Action: restart the system restart the system
missing ProbeGeneral data - wrong ProbeID. Action: restart the system restart the system
missing ProbeScanFunc - wrong ProbeScanFuncID. Action: restart the system restart the system
Mode not implemented now restart the system
MotionColor-DSC 1st:Memory allocation failed restart the system
MotionColor-DSC 2nd:Memory allocation failed restart the system
Motion-DSC:Memory allocation failed restart the system
No Cassette in drive! Put cassette into drive of VCR. Put cassette into the VCR
Error Messages Actions