GE HEALTHCAREDRAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-85
2.) Lift up the OPIO Assy. And then, disconnect 3 cables from OPIO ASSY.
3.) Separate the cable that connects with HUB Board.
4.) Remove 12 screws, and separate HUB Board..
Figure 8-101 Disconnecting 3 cables from OPIO ASSY
Figure 8-102 Disconnecting all cables from HUB Board
Figure 8-103 Removing 12 screws