EasyManua.ls Logo

GE Voluson P8 - Page 515

GE Voluson P8
553 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARERAFT VOLUSON® P8 / VOLUSON® P6
DIRECTION 5459672-100, R
EVISION 6 DRAFT (JANUARY 17, 2013) PROPRIETARY SERVICE MANUAL
10-4 Section 10-3 - Maintenance Task Schedule
Endocavity Probe
Leakage
Current Checks
Also after corrective
maintenance or as required by
your facilities QA program.
Measurement Accuracy
Checks
Also after corrective
maintenance or as required by
your facilities QA program.
Probe/
Phantom(optional)
Checks
Check axial and lateral resolution (see Basic
User Manual Technical specifications).
Check Gain and TGC changes, vary the
focus and check reaction on screen.
Also after corrective
maintenance or as required by
your facilities QA program.
Functional Checks of all
probes Section 10-6-2
on page 10-9
Also after corrective
maintenance or as required by
your facilities QA program.
Table 10-2 Customer Care Schedule
Item Service at Indicated Time D W M A Notes

Table of Contents

Other manuals for GE Voluson P8

Related product manuals