GMC-I Messtechnik GmbH 29
Protection Class I Devices
with Terminals for Applied Parts
Measurement type PE(TS) - P1
– DUT mains plug to test socket
– Test probe P1 to P1 terminals
Schematic Diagram
Insulation resistance is measured between protective conductor
terminal PE and external, short-circuited applied parts which can
be contacted with test probe P1.
Wiring Diagram
Protection Class I Devices
with
Outputs for Safety Extra-Low Voltage
Measurement type PE(TS) - P1
– DUT mains plug to test socket
– Test probe P1 to P1 terminals
Schematic Diagram
Insulation resistance is measured between the PE terminal and
the safety extra-low voltage outputs, which must be contacted
one after the other with probe P1.
Wiring Diagram
Protection Class I Devices
with Exposed Conductive Parts
Measurement type LN(TS) - P1//PE(TS)
– DUT mains plug to test socket
– Test probe P1 to P1 terminals
Schematic Diagram
Insulation resistance is measured between short-circuited mains
terminals (L-N) and external conductive parts which can be con-
tacted with test probe P1 and are not connected to the housing,
as well as protective conductor terminal PE at the housing.
Wiring Diagram
Setting Measuring Parameters for RISO
Measuring
Parameter
Meaning
Measurement Type,
Suitable for
DUT Connection via
LN(TS)-PE(TS) PC I: Testing is conducted be-
tween short-circuited LN mains
terminals at the test socket and
the DUT’s PE terminal
Test socket, EL1, VL2E,
AT3 adapter (AT3-IIIE, AT3-IIS,
AT3-I IS32), AT16DI/AT32DI,
CEE adapter
LN(TS)-P1 Testing is conducted between
short-circuited LN mains termi-
nals at the test socket and test
probe P1.
Test socket, VL2E,
AT3 adapter (AT3-IIIE, AT3-IIS,
AT3-I IS32), AT16DI/AT32DI
P1 – P2
SECUTEST PRO or feature H01:
2-pole measurement between
test probes 1 and 2
(see section 6.6)
No connection (PC3)
PE(mains)-P1
Cable test: Testing is conducted
between the ground terminal at the
mains and test probe P1.
Permanent connection
PE(TS)-P1
Testing is conducted between the
PE terminal at the test socket and
test probe P1.
Test socket
LN(TS)-P1 //
PE(TS)
Testing is conducted between
short-circuited LN mains termi-
nals at the test socket
and test
probe P1, including PE at the test
socket.
Test socket, VL2E,
AT3 adapter (AT3-IIIE, AT3-IIS,
AT3-I IS32), AT16DI/AT32DI
UISO(set)
> 50 ... < 500 V Variable test voltage can be entered with the numeric
keypad

Table of Contents

Other manuals for Gossen MetraWatt SECUTEST PRO

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gossen MetraWatt SECUTEST PRO and is the answer not in the manual?

Gossen MetraWatt SECUTEST PRO Specifications

General IconGeneral
BrandGossen MetraWatt
ModelSECUTEST PRO
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals