GMC-I Messtechnik GmbH 5
1 Applications
1.1 Table: Types of DUTs – Tests – Standards
Attention!
!
The test instrument may not be used for measurements
within electrical systems!
1.2 Table: Single Measurements and Regulations
Key
Specified test
Testing after Repairs
/ Periodic Testing
DUTs
in accordance with
the following
standards
DIN EN 62638, draft
DIN VDE 0701-0702
IEC 62353:2007
DIN EN 62353:2008
(VDE 0751-1)
IEC 60974-4:2010
DIN EN 60974-4:2011
VDE 0544-4:2011
Electric devices
Work devices
Mains operated electronic devices
Hand-held electric tools
Extension cords
Household appliances
Data processing devices
Electrical medical
devices, applied parts
Welding units
Single measurements
per regulation
DIN EN 62638, draft
DIN VDE 0701-0702:2008
IEC 62353:2007
DIN EN 62353:2008
(VDE 0751-1)
IEC 60974-4:2010
DIN EN 60974-4:2011
VDE 0544-4:2011
Protective conductor resistance •••
Insulation resistance •••
Protective conductor current
Primary leakage current
Device leakage current
Touch current ••
Current from welding circuits
Patient leakage current
Leakage current from the applied part
Test methods
Alternative measurement method Equivalent (device)
leakage current)
••
Differential current measuring method
•••
Direct measuring method •••