GMC-I Messtechnik GmbH 65
12 Characteristic Values
1
Specified values are only valid for the display at the test instrument. Data
transmitted via the USB port may deviate from these values.
2
Known as equivalent leakage current or equivalent patient leakage current
from previous standards
3
Protective conductor current, touch current, device leakage current, patient
leakage current
4
Protective conductor current, touch current, device leakage current
5
Only with SECUTEST BASE10 (feature G01) or SECUTEST PRO
6
Only with SECUTEST PRO (feature I01)
7
Measurement types IPE_clamp and IG_clamp
8
Measurement type IPE_AT3 adapter and IG_AT3 adapter
9
The upper range limit depends on the selected test voltage.
Key: rdg. = reading (measured value), d = digit(s)
Test Times, Automated Sequence
Test times (“measurement duration” parameter) can be set sepa-
rately for each rotary switch position during configuration of the
sequence parameters. Test times are neither tested nor calibrated.
Emergency Shutdown During Leakage Current Measurement
As of 10 mA of differential current (can also be set to 30 mA),
automatic shutdown ensues within 100 ms. This shutdown does
not take place during leakage current measurement with clamp
meter or adapter.
Func-
tion
Measured
Quantity
Display Range /
Nominal Range of
Use
Reso-
lution
Nominal
Voltage
U
N
Open-
Circuit
Voltage
U
0
Nom.
Current
I
N
Short-
Circuit
Current
I
K
Int.
Resis-
tance
R
I
Ref.
Resis-
tance
R
REF
Measuring
Uncertainty
1
Intrinsic Error
1
Overload
Capacity
Value Time
Tests, 62638 (DIN VDE 0701-0702) / IEC 62353 (VDE 0751)
Protective
conductor
resistance
R
PE
000 999 mΩ 1mΩ
< 24 V
AC or DC
> 200 mA
AC / DC
> 10 A AC
5
——
±(15% rdg.+ 10 d)
>10d
> 10.0 Ω:
±(10% rdg.+ 10 d)
±
(10% rdg.+ 10 d)
>10d
264 V
250 mA
16 A
5)
Cont.
1.00 … 9.99 Ω 10 mΩ
10.0 30.0 Ω
100 m
Ω
Insulation
resistance
9
RISo
10 999 kΩ 1kΩ
50 500
VDC
1.0 • U
N
1.5 • U
N
> 1mA < 2 mA
±(5% rdg.+ 4 d)
>10d
±
(2.5% rdg.+2 d)
> 10 d
264 V Cont.
1.00 9.99 MΩ 10 kΩ
10.0 99.9 MΩ 100 kΩ
20 MΩ:
±(10% rdg.+ 8 d)
20 MΩ:
±
(5% rdg.+4 d)
100 300 MΩ 1MΩ
Leakage current
alternative
measurement
2
IPE, IB, IG, IA
0.0 ... 99 μA1μA
50 ...
250 V~
20/+10%
—<1.5mA
>150k
Ω
1kΩ
±10 Ω
±
(5% rdg.+ 4 d) > 10 d
> 15 mA:
±(10% rdg.+ 8 d)
±
(2% rdg.+2 d) > 10 d
> 15 mA:
±
(5% rdg.+ 4 d)
264 V Cont.
100 ... 999 μA1μA
1.00 ... 9.99 mA 10 μA
10.0 ... 30.0 mA 100 μA
Leakage current
direct
measurement
3
IPE, IB, IG, IA, IP
Only Ip: 0.0 ...
99.9 μA
100 nA
——
1kΩ
±10 Ω
±(5% rdg.+ 4 d)
> 10 d
±
(2.5% rdg.+2 d)
> 10 d
264 V Cont.
0.0 ... 99 μA1μA
100 ... 999 μA1μA
1.00 ... 9.99 mA 10 μA
10.0 ... 30.0 mA 100 μA
Leakage current
differential
current
measurement
4
IPE, IB, IG
0 ... 99 μA1μA
——
1kΩ
±10 Ω
±(5% rdg.+ 4 d)
> 10 d
±
(2.5% rdg.+2 d)
>10d
264 V Cont.
100 ... 999 μA1μA
1.00 ... 9.99 mA 10 μA
10.0 ... 30.0 mA 100 μA
Function test
Line voltage
U
L–N
100.0 240.0 V~ 0.1 V
±
(2 % rdg.+2 d)
264 V Cont.
Load current I
L
0 16.00 A
RMS
10 mA
±
(2 % rdg.+2 d)
16 A Cont.
Active power P
0 3700 W 1 W
±
(5 % rdg.+10 d)
>20d
264 V Cont.
20 A 10 min
Apparent power S
0 4000 VA 1 VA Calculated value, U
L–N
• I
V
±
(5 % rdg.+10 d)
>20d
Power factor PF
with sinusoidal
waveform: cosϕ
0.00 1.00 0.01 Calculated value, P / S, display > 10 W ±(10 % rdg.+5 d)
Voltage Measurement
Probe voltage
(probe P1 to PE)
, and
0.0 99.9 V
100 ... 250 V
100 mV
1 V
——
3 MΩ
——
±
(2 %
rdg
.+2 d)
300 V Cont.
Measuring
voltage
(V–COM
sockets
6
, and
1 MΩ
±
(2 %
rdg
.+2 d)
> 45 Hz ... 65 Hz
±
(2 %
rdg
.+5 d)
> 65 Hz ... 10 kHz
±
(5 %
rdg
.+5 d)
> 10 kHz ... 20 kHz
t
A
PRCD
Time to trip
0.1 ... 999 ms 0.1 ms 30 mA
±
5 ms
I
Clamp
Current via
WZ12C current/
voltage trans-
former clamp
[1 mA:1 mV]
(V–COM sockets
67
)
1 ... 99 mA
1mA
(1 mV)
——
±(2 % rdg.+2 d)
>10d
20 Hz ... 20 kHz
without clamp
253 V Cont.
0.1 ... 0.99 A
0.01 A
(10 mV)
1.0 ... 9.9 A
0.1 A
(100 mV)
10 ... 15 A
1A
(1 V)
I
Leak-
age
Leakage current
via AT3-IIIE
adapter Z745S
6 8
0.00 ... 0.99 mA
0.01 mA
——
±(2 % rdg.+2 d)
>10d
without adapter
253 V Cont.
1.0 ... 9.9 mA
0.1 mA
10 ... 20 mA
1mA
Temp
Temperature
with Pt100
- 200,0 ...
+850.0 °C
0.1 °C— <20V 1.1mA ±(2% rdg.+1 °C) 10 V Cont.
Temperature
with Pt1000
- 150,0 ...
+850.0 °C