66 GMC-I Messtechnik GmbH
Influencing Quantities and Influence Error
Reference Ranges
Line voltage 230 V AC ±0.2%
Line frequency 50 Hz ±2Hz
Waveform
Sine (deviation between effective and rectified value < 0.5%)
Ambient temperature +23 °C ±2K
Relative humidity 40 … 60%
Load resistance Linear
Nominal Ranges of Use
Nominal line voltage 100 V … 240 V AC
Nom. line frequency 50 Hz ... 400 Hz
Line voltage waveform Sinusoidal
Tem perat ure 0 °C … +50°C
Ambient Conditions
Storage temperature – 20 °C … +60°C
Operating temperature– 5 °C … +40°C
Accuracy range 0 °C … +40°C
Relative humidity Max. 75%, no condensation allowed
Elevation Max. 2000 m
Deployment
Indoors, except within specified ambient
conditions
Power Supply
Supply network TN, TT or IT
Line voltage 100 V … 240 V AC
Line frequency 50 Hz ... 400 Hz
Power consumption 200 mA test: approx. 32 VA
10 A test: approx. 105 VA
For function test Continuous max. 3600 VA, power is con-
ducted through the instrument only,
switching capacity: ≤ 16 A, ohmic load
Electrical Safety
Protection class I per IEC 61010-1/EN 61010-1/
VDE 0411-1
Nominal voltage 230 V
Test voltage 2.3 kV AC 50 Hz or 3.3 kV DC
(mains circuit / test socket to mains PE ter-
minal, USB, finger contact, test probe P1,
test socket)
Measuring category 250 V CAT II
Pollution degree 2
Safety shutdown At DUT differential current of > 10 mA,
shutdown time: < 100 ms,
can also be set to > 30 mA
with following probe current during:
– Leakage current meas.: > 10 mA~/
<5ms
– Protective conductor resistance meas.:
>250mA~/<1ms
Fuse links Mains fuses: 2 ea. FF 500 V / 16
A
Probe fuse: M 250 V / 250 mA
SECUTEST BASE10/PRO:one additional
FF 500 V / 16 A FF
Electromagnetic Compatibility
Product Standard DIN EN 61326-1
USB Data Interface
Type USB slave for connection to a PC
Typ e 2 ea. USB master,
for data entry devices with HID
interface
(e.g. keyboard, barcode/RFID scanner),
for USB flash drive for data backup,
for USB flash drive for saving
reports as BMP files
for printers
Mechanical Design
Display 4.3" multi-display (9.7 x 5.5 cm), backlit,
480 x 272 pixels
at 24 bit color depth (true color)
Dimensions W x H x D: 295 x 145 x 150 mm
Height with handle: 170 mm
Weight Approx. 2.5 kg
Protection Housing: IP 40
Test socket: IP 20 per DIN VDE 0470, part
1/EN 60529
Table Excerpt Regarding Significance of IP Codes
Influencing Quantity /
Sphere of Influence
Designation
per
DIN VDE
0404
Influence error
± …% rdg.
Change of position E1 —
Change to test equipment supply
voltage
E2 2.5
Temperature fluctuation
E3
Specified influence error valid
starting with temperature changes
as of 10 K:
0 ... 40
°C2.5
Amount of current at DUT E4 2.5
Low frequency magnetic fields E5 2.5
DUT impedance E6 2.5
Capacitance during insulation
measurement
E7 2.5
Waveform of measured current
E8
49 … 51 Hz
2 with capacitive load (for equiva-
lent leakage current)
45 … 100 Hz 1 (for touch current)
2.5 for all other measuring ranges
Interference
emission
Class
EN 55011 B
Interference
immunity
Test Value Evaluation criterion
EN 61000-4-2 Contact/atmos. – 4 kV/8 kV A
EN 61000-4-3 3 V/m or 1 V/m A
EN 61000-4-4 1 kV B
EN 61000-4-5 1 kV or 2 kV A
EN 61000-4-6 3 V/m A
EN 61000-4-11 0.5/1/25 periods A
250 periods C
IP XY
(1
st
digit X)
Protection Against
Foreign Object Ingress
IP XY
(2
nd
digit Y)
Protection Against
Penetration by Water
2 ≥ 12.5 mm dia. 0 Not protected
4 ≥ 1.0 mm dia. 0 Not protected